Welcome to Pascal Control | Malaysia Valve Supplier
LUGB & LUCB Vortex flowmeter work on the principle of generated vortex and relation between vortex and flow by theory of Karman and Strouhal, which specialize in measurement of steam, gas and liquid of lower gas and vapor even turbid liquid including micro grain and impurity. Vortex Flowmeter is on the principle of Karman Street, to measure liquid, sides with opposite directions of rotation, Vortices frequency is directly proportional to medium velocity. Through numbers of vortices that is measured by sensor head, medium velocity is calculated, plus flowmeter diameter, final volume flow come out.
Application Range |
|
Measured Value | |
Primary Measured Value | Flow Rate |
Secondary Measured Value | Volume flow (Pressure and Temperature is available) |
Temperature | |
Process Temperature | T1 Level: -20℃…+100℃ |
T2 Level: -20℃…+250℃ | |
T3 Level: -20℃…+350℃ | |
T4 Level: -20℃…+420℃ | |
Ambient Temperature | -10℃…+50℃ |
Pressure | |
EN 1092-1 | DN200…DN300: PN10 |
DN100…DN200: PN16 | |
DN15…DN80: PN25 | |
Other pressure in request | |
ASME B 16.5 | 1/2”…8”: 150 Ib RF |
Other pressure in request | |
JIS | 1/2”…8”: 10K |
Other pressure in request | |
Reference Condition | Flow conditions similar to EN 29104 |
Medium: Water/Gas/Steam | |
Temperature: -10℃…+30℃ | |
Inlet Section: ≥10DN | |
Accuracy | For Liquid: ±1.0% of rate |
For Steam and Gas:±1.5% of rate | |
Body Material | SS304 |
SS306 | |
Converter Material | Standard: Polyurethane coated die-cast aluminum |
Model | Suffix Code | Description | |||||||
LUGB- | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Vortex Flowmeter |
Fluid | L | Liquid | |||||||
G | Gas/Air | ||||||||
S | Steam | ||||||||
Diameter | XXX | Stand for diameter 015: DN 15; 050: DN 50; 100: DN 100; 300: DN 300 |
|||||||
Structure | S | Compact type | |||||||
L | Remote type | ||||||||
Converter Type | C | Fluid: liquid; 24V DC; 4-20mA/Pulse output; Digital display; Ex | |||||||
V | 24V DC; 4-20mA/Pulse output ( V type is only for Gas/Steam application ); Digital display; Ex | ||||||||
D | 24V DC; 4-20mA/Pulse output; Temperature & Pressure compensation; Digital display; Ex | ||||||||
DB | 24V DC; 4-20mA/Pulse output; Temperature & Pressure compensation; Digital display; Ex | ||||||||
DA | 24V DC; 4-20mA/Pulse output; Temperature & Pressure compensation; Digital display; Ex | ||||||||
DV | 24V DC; 4-20mA/Pulse output; Temperature & Pressure compensation; Digital display; Ex | ||||||||
Notice: |
|
||||||||
Body Material | S4 | SS304 | |||||||
S6 | SS306 | ||||||||
Explosion Proof | BT | ExdIIBT6 | |||||||
CT | ExibIICT4 | ||||||||
NA | No explosion proof | ||||||||
Connection | WAF | Wafer connection | |||||||
DXX | D16: DIN PN 16 Flange; D25: DIN PN 25 Flange… | ||||||||
AXX | A15: ANSI 150# Flange; A30: ANSI 300# Flange… | ||||||||
JXX | J10: JIS 10K Flange; J20: JIS 20K Flange… | ||||||||
Temperature Rating | T1 | -20℃…+100℃ | |||||||
T2 | -20℃…+250℃ | ||||||||
T3 | -20℃…+350℃ | ||||||||
T4 | -20℃…+420℃ |